7 appendix, 1 technical data, 732 ic detector – Metrohm 732 IC Detector User Manual
Page 196: 733 ic separation center, 7appendix
7.1 Technical data
732 IC Detector / 733 IC Separation Center
189
7
Appendix
7.1
Technical data
7.1.1
732 IC Detector
Conductivity measurement
Measurement ranges
100, 200, 500, 1000, 2000, 5000, 10 000
µ
S/cm
Full scale
0.05
…
10 000
µ
S/cm
Temperature correction
Automatic correction of the measured conductiv-
ity to reference temperature 20°C with the
adjustable temperature coefficients
(1.5 %/°C or 2.5 %/°C)
Accuracy, absolute value
< 4 % of full-scale value for all ranges under the
following reference conditions:
Adjustment with reference resistance to 1.000 V
(range 1 mS/cm, cell constant 16.7 /cm, temp.
coeff. 2.5 %/°C, frequency 5 kHz, room tempera-
ture) to accuracy <
±
2 mV, measurement
accuracy < 0.5 % after 30 min warm-up time
Linearity
Up to
±
150 % of each full-scale range
Deviations < 0.5 % of full-scale range
Temperature dependence
Typically 25 ppm/°C
Dependence of cell const..
<
±
1 % over entire adjustment range
Drift (electronic)
< 0.0013 % of selected measurement range
/h/°C
Noise (electronic)
Typically < 0.0003 % of selected meas. range
Damping
2-stage damping (Bessel 4
th
order)
"
damping
=
off
"
0.25 s (10
…
90 %)
"
damping
=
on
"
2.00 s (10
…
90 %)
Auto-zero
Function
Automatic zero setting (electronic background
compensation) over entire selectable measure-
ment range
Initiation
Manually or externally (RS232, remote)
Maximum error
±
0.6 % of full-scale range
Linear range
±
150 % of full-scale range
Marker
Signal
Ca. 10 % of full-scale range
Initiation
Manual or externally (RS232, remote)