6.1 RS232 interfaces

732 IC Detector / 733 IC Separation Center

165

6.1.5

Error messages

Error messages ‘

EXXX

’ are appended to the status message and sepa-

rated from this by a ‘

;

’.

Error

Meaning

Exit/Corrective action

E28

Wrong object call-up.

Correct path.

E29

Wrong value.

Enter correct value or new path.

E30

Wrong trigger.

Enter correct trigger or new path.

E36

RS232 receive error; parity.

, set same parity for both devices.

E37

RS232 receive error; stop bit.

, set same stop bit for both devices.

E38

RS232 receive error; overflow (at least 1
character could not be read).

, set same baud rate for both
devices, restart sender.

E39

RS232 receive error; overflow of internal receive
buffer (>82 characters).

.

E40

RS232 send error; DSR=OFF. Handshake not
answered for longer than 1 s.

, check receiver
(switched on and ready?).

E41

RS232 send error; DCD=ON.
Handshake not answered for longer than 1 s.

, check receiver
(switched on and ready?).

E42

RS232 send error; CTS=OFF. Handshake not
answered for more than 1 s.

, check receiver
(switched on and ready?).

E43

RS232 send error; transmission of the IC
Detector was interrupted with XOFF for at least
3 s.

or send XON.

E44

RS232 send error ; the RS parameters are no
longer the same for both devices.

, reset RS parameters for both
devices.

E45

RS232 send error; the receive buffer of the IC
Detector contains an incomplete string (L

FF

missing), the transmission is thus blocked.

or send L

FF

.

E50

E59

Error in IO test.

, inform Metrohm Service.

E60

E89

Error in RS232 test.

, check connection between the RS
interfaces.

E200

Instrument adjustment invalid.

Inform Metrohm Service.

E202

Wrong entry via RS232 remote control.

Enter correct value, correct trigger or new
path.

E240

Error in storage of the EEPROM blocks.

, re-enter all values.