Metrohm 732 IC Detector User Manual
Page 172
![background image](/manuals/586520/172/background.png)
6.1 RS232 interfaces
732 IC Detector / 733 IC Separation Center
165
6.1.5
Error messages
Error messages ‘
EXXX
’ are appended to the status message and sepa-
rated from this by a ‘
;
’.
Error
Meaning
Exit/Corrective action
E28
Wrong object call-up.
Correct path.
E29
Wrong value.
Enter correct value or new path.
E30
Wrong trigger.
Enter correct trigger or new path.
E36
RS232 receive error; parity.
E37
RS232 receive error; stop bit.
E38
RS232 receive error; overflow (at least 1
character could not be read).
devices, restart sender.
E39
RS232 receive error; overflow of internal receive
buffer (>82 characters).
E40
RS232 send error; DSR=OFF. Handshake not
answered for longer than 1 s.
(switched on and ready?).
E41
RS232 send error; DCD=ON.
Handshake not answered for longer than 1 s.
(switched on and ready?).
E42
RS232 send error; CTS=OFF. Handshake not
answered for more than 1 s.
(switched on and ready?).
E43
RS232 send error; transmission of the IC
Detector was interrupted with XOFF for at least
3 s.
E44
RS232 send error ; the RS parameters are no
longer the same for both devices.
devices.
E45
RS232 send error; the receive buffer of the IC
Detector contains an incomplete string (L
FF
missing), the transmission is thus blocked.
FF
.
E50
…
E59
Error in IO test.
E60
…
E89
Error in RS232 test.
interfaces.
E200
Instrument adjustment invalid.
Inform Metrohm Service.
E202
Wrong entry via RS232 remote control.
Enter correct value, correct trigger or new
path.
E240
Error in storage of the EEPROM blocks.