beautypg.com

Function, Command format, Relevant query command – Spectra Precision ProFlex 800 Reference Manual User Manual

Page 495: See also

background image

481

Set Command Library

$PASHS,ETH

TLT,CMD: Defining the Trigger String Used to Query the Tiltmeter

Function

This command is used to define the character string that will
query the tiltmeter. The command also specifies the ID of the
receiver port used to communicate with the tiltmeter.

The trigger string is in the form “*xxxxxx” and the default one
is *0100XY.

Command Format

Syntax

$PASHS,TLT,CMD,c1,s2[*cc]

Parameters

Example

Setting trigger string to “*0100XY”, tiltmeter connected to
port F:

$PASHS,TLT,CMD,F,0100XY*19

Relevant Query

Command

$PASHQ,TLT

See Also

$PASHS,TLT,INIT

$PASHS,TLT,INTVL

$PASHS,OUT,x,TLT

Parameter

Description

Range

c1

Receiver serial port connected to the tiltmeter. A, B, F

s2

Trigger string (not including the leading “*”
character)

20 characters max.

*cc

Optional checksum

*00-*FF