51
Voyager M3x Analyzer
Teledyne LeCroy Corporation
Note:
USB 2.0 Link LEDs operate only while USB 2.0 Recording or Real‐Time Statistics (RTS) is running.
USB 3.0 LEDs always operate, unless USB 3.0 has been disabled in the Recording Options
General Tab.
WARNING:
Do not open the Voyager M3x enclosure. No operator serviceable parts are
inside. Refer servicing to Teledyne LeCroy customer care.
2.4.4
Specifications
The Analyzer has the following specifications.
Power Requirements
12 V DC, 5 amps.
In
SMA external trigger input. Note: Edge
detected. (Rising edge only) Voltage required:
Signal needs to be > 800mV to see a logic "1",
Signal needs to be < 400mV to see a logic "0" .
Maximum value of the external input signal
which can be input is 5 V. Minimum value is 0
V.
Out
SMA external trigger output.
Pulse Width = 280 ns. Pulse Voltage = 3.3
Volts into 1 MegOhms. Pulse Voltage = 2.7
Volts into 50 Ohms. Pulse is positive going.
Rear Panel:
Sync In
For use in Daisy chaining and Crosssync
scenarios.
Sync Out
For use in Daisy chaining and Crosssync
scenarios.
ETHERNET
Gigabit Ethernet for connection to host
machine.
USB
USB 3.0 Super Speed Standard B Connector to
host machine.
DC IN
12V, 5A power supply connector.
Power Switch
(0/1) Off/On