4.4 Using the Sequence Function

APS-1102 Programmable AC/DC Power Source

4-33

Table 4-10. Setting Items in PROGRAM Screen (3/3)

Setting Name

Item

Operation

STEP TERM

Step end

Sets operation after end of step.
Select one of the following three options.
 CONT: Sequence continuation

 STOP:

Change to Idle mode

 HOLD: Change to Hold mode

JUMP STEP

Jump step

Sets the step number to jump to after the end of the
previous step.
 0:

Continue to next step number

 1 to 255: Jump to specified step number

JUMP NUM

Jump times

Sets the number of times to execute the jump operation
specified by JUMP STEP.
 0:

Unlimited jumps

 1 to 999: Specified number of jumps

BRANCH0 Branch

step

0

Sets the step number to jump to when trigger input has
occurred.
Two types of branch steps can be set: BRANCH0 and
BRANCH1.
[Command example]

When a STOP is set after a small number of sweeps at the
jump destination step, it enables uses where there is a
safe region for saving data if necessary.

 0:

Branch disabled

 1 to 255: Jump to specified step number

BRANCH1 Branch

step

1


The operation steps for each of the setting items in the PROGRAM screen are described below.

Operation step <1>
The settings for step number (STEP), step time (TIME), DC voltage (DC VOLT, AC+DC mode only),
AC voltage (AC VOLT), AC frequency (FREQ), phase (PHS), step end phase (STOP PHS), jump step
(JUMP STEP), jump times (JUMP NUM), and branch (BRANCH0/BRANCH1) are made by entering
numerical values for various items in the order described below.

The example described below is when entering the setting for step time (TIME).