6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-31

STATus:OPERation:NTRansition
STATus:OPERation:PTRansition
This sets the operation transition filter (OPTF). The setting ranges from 0 to 20482.
 When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
 When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1.
 When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
 When the NTR and PTR filters are both set to 0, changes in OPCR are not propagated to OPER.

Syntax:

STATus:OPERation:NTRansition

STATus:OPERation:PTRansition

Parameters:

0 to 20482

Command example:

STAT:OPER:NTR 16384

STAT:OPER:PTR

0

Query example:

STAT:OPER:NTR?

STAT:OPER:PTR?

Response format:

Response example:

16384 (when NTR is queried)

0 (when PTR is queried)

STATus:WARNing:CONDition?
The warning condition register (WRCR) value is returned. The response ranges from 0 to 31999.

Syntax:

STATus:WARNing:CONDition?

Parameters:

None

Query example:

STAT:WARN:COND?

Response format:

Response

example: 16384