4.8 Using Memory Functions
APS-1102 Programmable AC/DC Power Source
4-53
4.7.2 Line synchronization
This function synchronizes the output frequency to the frequency of the AC power source line. The
output frequency can be synchronized to power source line frequency (50 Hz/60 Hz).
Operation steps
Select
in the SET menu to select external synchronization mode, then set the sync signal source
to line synchronization (“LINE”).
1. Perform the same operations as were described above for steps 2 to 5 in “4.7.1 External signal
synchronization” to select external synchronization mode, after which the external sync signal
source selection box will be displayed.
ENTER
CANCEL
Selection box
2. Press the
or
key to move the cursor to “LINE”, the press the
ENTER
key.
This sets the external sync signal source to “LINE”.
ENTER
CANCEL
4.8 Using Memory Functions
Various settings are retained in memory with battery backup, and the settings when the power source was
last turned off are restored at power-on.
Among these various settings, the basic settings (output mode, output range, DC settings, AC settings,
output current limiter, and setting range limit) can be stored in store/recall memory No. 1 to No. 30, and
these stored settings can be recalled.
The recall operation for store/recall memory can be executed only when output is off.