7.2 Error Messages and Responses

APS-1102 Programmable AC/DC Power Source

7-9

Table 7-5. Panel Operation Errors (2/2)

Error Message

Cause and Action Required

UNDER BUSY STATE

Cannot set (execute) due to BUSY status. Wait until the BUSY display
disappears, then try operation again.
(

See “3.4.3 Setting output voltage range”.)

UNDER REMOTE STATE

Cannot set (execute) due to remote status. Set local mode.
(

See “5.5.3 Remote”.)

UNDER LOCAL STATE

Cannot set (execute) due to local mode. (Remote mode only)
(

See “4.5 Control Using External Control I/O Connector”.)

(

See “6 USB INTERFACE”.)

UNDER KEYLOCK

Cannot set (execute) due to keylock status. Cancel keylock.
(

See “5.6.1 Keylock”.)

OUT OF LIMITER

Cannot set (execute) because range limit is exceeded. (Remote mode
only)
(

See “4.5 Control Using External Control I/O Connector”.)

(

See “6 USB INTERFACE”.)

COMMAND ERROR

This is the USB interface command error. Check the command.
(Remote mode only)
(

See “6.6 Error Message List”.)

EXECUTION ERROR

Execution errors other than the above.