beautypg.com

1 output – GW Instek APS-1102 User Manual User Manual

Page 250

background image

9.1 Output

APS-1102 Programmable AC/DC Power Source

9-2


Unless otherwise noted, the following conditions are adopted:

Power input:

200 V AC, 50 Hz

Output mode:

AC-INT (alternate current - internal signal source) mode

Output frequency:

50 Hz

Output voltage:

100 Vrms (200 Vrms when in 200 V range)

Output waveform:

Sine wave

Load

100 V range: 10 Ω (power factor 1, nominal value)

200

V

range:

40

Ω (power factor 1, nominal value)

Output terminals:

Terminal block on rear panel

9.1 Output

Output mode
All eight modes (combination of operation mode + signal source mode)

AC-INT mode (alternate current - internal signal source)
AC-EXT mode (alternate current - external signal source)
AC-ADD mode (alternate current - internal + external source)
AC-SYNC mode (alternate current - external synchronization)
AC+DC-INT mode (direct current - internal signal source)
AC+DC-EXT mode (direct current - external signal source)
AC+DC-ADD mode (direct current - internal + external source)
AC+DC-SYNC mode (direct current - external synchronization)

Output voltage ranges
100 V range and 200 V range

Maximum output power
750 VA (AC) / 750 W (DC)

Power supply input conditions: 100 V AC to 180 V AC input

(hereinafter referred to as “AC 100 V input system”)

Output voltage conditions:

130 Vrms (AC) /

183 V (DC) or lower (100 V range)

260

Vrms

(AC)

/

366 V (DC) or lower (200 V range)


1000 VA (AC) / 1000 W (DC)

Power supply input conditions: 180 V AC to 250 V AC input

(hereinafter referred to as “AC 200 V input system”)

Output voltage conditions:

135 Vrms (AC) /

190 V (DC) or lower (100 V range)

270

Vrms

(AC)

/

380 V (DC) or lower (200 V range)