4 list of functions – GW Instek APS-1102 User Manual User Manual
Page 18
1.4 List of Functions
APS-1102 Programmable AC/DC Power Source
1-4
1.4 List of Functions
The main functions of the APS-1102 are listed below.
Table 1-1. List of Functions (1/2)
Function
Description
Output
system
Output mode
Total of 8 modes
(combinations of operation mode and signal source mode)
Alternate current – internal signal source
(AC-INT mode)
Alternate current – external signal source
(AC-EXT mode)
Alternate current – internal + external signal sources
(AC-ADD mode)
Alternate current – external synchronization
(AC-SYNC mode)
Direct current – internal signal source
(AC+DC-INT mode)
Direct current – external signal source
(AC+DC-EXT mode)
Direct current – internal + external signal sources
(AC+DC-INT mode)
Direct current – external synchronization
(AC+DC-SYNC mode)
Output on/off switch
Output voltage range
100 V range and 200 V range
Output voltage waveform
(not including external signal
source mode)
Sine wave, square wave, arbitrary waveform (16 types)
Output current limiter
Output peak current limiter function and output average
current limiter function (variable limit values)
Setting range limiter
(not including external signal
source mode)
Setting range limiter function for output voltage and
output frequency
Internal mode (AC-INT, AC+DC-INT) and internal +
external mode (AC-ADD, AC+DC-ADD) only
Setting range limiter function for output voltage
External synchronization mode (AC-SYNC,
AC+DC-SYNC)
Sequential output
Output parameters: jump or sweep.
Synchronized output
Can be synchronized with external sync signal or line
frequency
Measure-
ment
function
Basic measurements
Voltage: RMS, average, peak
Current: RMS, average, peak, peak hold
Power: Effective, reactive, apparent
Other
Synchronization frequency: Only in SYNC mode
Load power factor
Load crest factor
Harmonic current: 50/60 Hz fundamental only, up to
40th