9.5 Measurement Functions
APS-1102 Programmable AC/DC Power Source
9-9
Output power measurement
Effective
power
Full
scale:
1200
W
Display
resolution:
1
W
Measurement accuracy: ±2% of the full scale (at 23 ±5°C)
(when output is 100 VA or above)
Apparent power:
Calculated as output voltage RMS value × output current RMS value
Full
scale:
1400
VA
Display
resolution:
1
VA
Measurement accuracy: ±2% of the full scale (at 23 ±5°C)
(when output is 100 VA or above)
Reactive power:
Calculated as
(apparent power)
2
– (effective power)
2
Full
scale:
1400
Var
Display
resolution:
1
Var
Measurement accuracy: ±2% of the full scale (at 23 ±5°C)
(when output is 100 VA or above)
Load power factor measurement: Calculated as effective power / apparent power
Measurement range:
0.00 to 1.00
Display resolution:
0.01
Load crest factor measurement: Calculated as (
positive peak current and negative
peak current
maximum value) / RMS current
Measurement range:
0.00 to 50.00
Display resolution:
0.01
Output harmonic current measurement
(AC - INT mode only, fundamental wave is 50 or 60 Hz only)
Measurement range:
To 40th harmonic of fundamental wave
Full scale:
15 Arms and 100%
Display resolution:
0.01 Arms and 0.1%
Measurement accuracy:
1% of full scale at RMS value current (at 23 ±5°C, to 20th harmonic)
±1.5% of full scale at RMS value current (at 23 ±5°C, from 20th to
40th)
Note: This measurement does not comply with the IEC standards.
External synchronization frequency measurement
(external synchronization mode only)
Measurement range:
38.0 to 525.0 Hz
Display resolution:
0.1 Hz
Measurement accuracy:
±0.2 Hz (at 23 ±5°C, 50 or 60 Hz)