beautypg.com

GW Instek APS-1102 User Manual User Manual

Page 95

background image

4.4 Using the Sequence Function

APS-1102 Programmable AC/DC Power Source

4-31

Table 4-8. Setting Items in PROGRAM Screen (1/3)

Setting Name

Item

Operation

STEP Step

number

Specifies the step number to be set.
Setting range: 1 to 255

TIME Step

time

Specifies the continuation time for the specified step.
Setting range:

0.0001 s to 999.9999 s

Setting resolution: 0.0001 s (= 0.1 ms)
 The time unit can be switched between “s” and “ms” via

the SYSTEM menu “TIME UNIT” setting.

See “5.6.5 Time unit setting”.

DC VOLT
(AC+DC-INT
only)

DC voltage value

Sets the DC voltage value for AC+DC mode.
The setting range is determined by the range setting used
when editing sequence data.

Setting range: 100 V range:

200.0 V to +200.0 V

200 V range:

400.0 V to +400.0 V

Setting resolution: 0.1 V

DC operation type

Sets the method used to change to the specified DC voltage
value.
Select one of the following three options.
 CONST: Output specified value
 KEEP:

Output value at end of previous step

 SWEEP: Start by outputting value at end of previous

step, then perform a linear sweep to output the
specified value at the end of the current step

AC VOLT

AC voltage value

Sets the AC voltage value.
The setting range is determined by the range setting entered
when editing the sequence data.
Setting range: 100 V range: 0.0 Vrms to 140.0 Vrms

0.0 Vp-p to 400.0 Vp-p

200 V range: 0.0 Vrms to 280.0 Vrms

0.0 V p-p to 800.0 Vp-p

Setting resolution: 0.1 Vrms/0.1 Vp-p
 The voltage value can be set as “Vp-p” only when

arbitrary waveform has been selected.

AC operation type

Sets the method used to change to the specified AC voltage
value.
Select one of the following three options.
 CONST: Output specified value
 KEEP:

Output value at end of previous step

 SWEEP: Start by outputting value at end of previous

step, then perform a linear sweep to output the
specified value at the end of the current step