4.4 Using the Sequence Function

APS-1102 Programmable AC/DC Power Source

4-23

The step transition parameters specify the next step and its timing.

Table 4-5. Step Transition Parameters

[Step time]

Specified in time

[Step end phase]

Sets step end phase as enabled or disabled.

[Step end]

The following parameters are provided to specify the action at the end of the
step.
 Continue sequence (CONT)

 Hold (HOLD)
 Idle (STOP)

[Jump step]

The following parameters are provided to specify the destination step when a
jump occurs after ending the current step.
 Jump to an arbitrary number step (1 to 255)
 Jump to the next number step (0)

[Jump times]

The following parameters set a loop for jumping to the step specified by [jump
step] the specified number of times.
The following parameters are provided to specify the jump times.
 Specified number (1 to 999)

 Unlimited (0)

[Branch step]

The following parameters are provided to specify the destination step for
branch control.
 Change to arbitrary step number (1 to 255)

 Continue step processing without changing (0)


When start is commanded during normal output while in idle mode, step 1 is executed, then the step jumps
according to the specified sequence.
If the [jump step] has not been specified (= 0), the step jumps according to the sequence of step numbers,
but when the [jump step] has been specified, the step jumps to the step specified by that parameter.
A step ends when the time specified by the [step time] has elapsed. If the [step end phase] has been
specified as enabled, the step ends when the [step end phase] value becomes (0

) after the [step time] has

elapsed.
If both [step end] and [jump step] have been specified, the [step end] specification takes priority.