beautypg.com

2 standard event status – GW Instek APS-1102 User Manual User Manual

Page 217

background image

6.5 Status System

APS-1102 Programmable AC/DC Power Source

6-45

6.5.2 Standard event status

The structure of the standard event status register is shown in Figure 6-5 below.

Power on (PON)

User request (URQ)

Command error (CME)

Execution error (EXE)

Device specific error (DDE)

Query error (QYE)

Request control (RQC)

Operation complete (OPC)

7
6
5
4
3
2
1
0

ESR (standard event register)

7
6
5
4
3
2
1
0

Logi

cal OR

ESE (standard event enable register)

Standard event status summary

Status byte (bit 5)

Figure 6-5. Standard Event Status Register


The definition of the standard event status register is listed in Table 6-14. Bits in the standard event
status register become valid when 1 is set to the standard event status enable register, and the ORed result
of the valid bits is reflected in the ESB bit of the status bit register.

The standard event status register can be read by an

ESR? query.

All of the bits are cleared when they are read by an *ESR? query, when a *CLS command is executed, or
when the power is turned on again (except that the PON bit is set to 1 when the power is turned on again).