5.5 MISC Menu


APS-1102 Programmable AC/DC Power Source

5-28

b) PROGRAM (sequence programming)

This “PROGRAM” screen is used to program sequences.
The items listed below in Table 5-20 to Table 5-22 are set for individual steps.
Sequence programming is stored automatically, and the data for AC mode with a 100 V range or a 200
V range is stored separately from the data for AC+DC mode with a 100 V range or a 200 V range.

See “4.4.3 Programming sequence”, for description of sequence programming operation

steps.

Table 5-20. PROGRAM Screen Items (1/3)

Setting Name

Item

Operation

STEP Step

number

This specifies the step number to be set.
Setting range: 1 to 255

TIME Step

time

Specifies the continuation time for the specified step.
Setting range:

0.0001 s to 999.9999 s

Setting resolution: 0.0001 s (= 0.1 ms)
 The time unit can be switched between “s” and “ms” via

the SYSTEM menu “TIME UNIT” setting.

See “5.6.5 Time unit setting”.

DC VOLT
(AC+DC-INT
only)

DC voltage value

Sets the DC voltage value for AC+DC mode.
The setting range is determined by the range setting used
when editing sequence data.

Setting range: 100 V range:

200.00 V to +200.00 V

200 V range:

400.00 V to +400.00 V

Setting resolution: 0.1 V

DC operation type

Sets the method used to change to the specified DC voltage
value.
Select one of the following three options.
 CONST: Output specified value

 KEEP: Output value at end of previous step

 SWEEP: Start by outputting value at end of previous

step, then perform a linear sweep to output the
specified value at the end of the current step

AC VOLT

AC voltage value

Sets the AC voltage value.
The setting range is determined by the range setting entered
when editing the sequence data.

Setting range: 100 V range: 0.0 Vrms to 140.0 Vrms

0.0 Vp-p to 400.0 Vp-p

200 V range: 0.0 Vrms to 280.0 Vrms

0.0 Vp-p to 800.0 Vp-p

Setting resolution: 0.1 Vrms/0.1 Vp-p

 The voltage value can be set as “Vp-p” only when

arbitrary waveform has been selected.

AC operation type

Sets the method used to change to the specified AC voltage
value.
Select one of the following three options.
 CONST: Output specified value

 KEEP: Output value at end of previous step
 SWEEP: Start by outputting value at end of previous

step, previous perform a linear sweep to output
the specified value at the end of the current step