6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-34

i) TRACe subsystem

TRACe:CATalog?
This returns a list of defined waveform names. This command returns a list of all waveform
names that can be selected by the user.

Syntax:

TRACe:CATalog?

Parameters:

None

Query example:

TRAC:CAT?

Response format:

[, ]

Response example:

SIN, SQU, ARB1, ARB2, ARB3, ARB4, ARB5, ARB6, ARB7, ARB8,

ARB9, ARB10, ARB11, ARB12, ARB13, ARB14, ARB15, ARB16

TRACe[:DATA]
This sends or receives arbitrary waveforms. When a waveform name (ARB1 to ARB16) is
selected as the first parameter, the subsequent parameters are used to send 2 bytes × 4096 words (=
8192 bytes) of waveform data in big endian order (one word of waveform data is sent in 2’s
complement format, higher byte first and then lower byte). As in the command example,
everything from the command to the data should be sent at one time.
However, the waveform data must be 4096 words.
The effective range for waveform data is

16384 to 16383. If a value of 16384 or below is input,

it is returned as

16384. If a value of 16383 or above is input, it is recognized as 16383.

Syntax:

TRACe[:DATA] ,

First parameter (waveform name)

Type

Range

ARB1

to

ARB16

Second parameter and afterward (binary block data)

Configuration #48192...

Command example:

TRAC ARB1, #48192...

Query example:

TRAC? ARB1

Response format:

Response