6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source
6-21
[SOURce:]SEQuence:TPARameter 
This sets the step transition parameters. 
The target sequence step must be set in advance by the SEQ:STEP. (
See
“[SOURce:]SEQuence:STEP” on page 6-22.) 
Use SYST:TUN (
See “ SYSTem:TUNit” on page 6-36) in advance to set the unit for the
sequence step execution time or query the unit. 
When in AC-EXT, AC-SYNC, AC+DC-EXT, or AC+DC-SYNC mode, the step transition 
parameter cannot be set. ([3, “Invalid in this mode”] error occurs.) 
Syntax:
[SOURce:]SEQuence:TPARameter 
 First parameter (step time [s]/[ms]) 
 
Type
Range
0.0001 to 999.9999 [s]
0.1 to 999999.9 [ms]
 Second parameter (step end phase setting mode) 
 
Type
Range
0 (disable) or 1 (enable)
Third parameter (step end phase [
])
Type
Range
0.0 to 359.9
 Fourth parameter (step termination) 
 
Type
Range
0 (continue), 1 (end), or 2 (hold)
 Fifth parameter (jump step number) 
 
Type
Range
0 (next step), 1 to 255
 Sixth parameter (jump times) 
 
Type
Range
0 (unlimited times), 1 to 999
 Seventh parameter (branch 0) 
 
Type
Range
0 (disabled), 1 to 999
 Eighth parameter (branch 1) 
 
Type
Range
0 (disabled), 1 to 999
 Command example: SEQ:TPAR 2, 1, 180, 0, 0, 1, 0, 0 
 Query 
example: SEQ:TPAR?
Response example: 2, 1, 180, 0, 0, 1, 0, 0