beautypg.com

2 current limiter, 3 setting range limits – GW Instek APS-1102 User Manual User Manual

Page 254

background image

9.2 Current Limiter

APS-1102 Programmable AC/DC Power Source

9-6

9.2 Current Limiter

Output peak current limiter
When limiter is operating, output voltage is clipped.

Positive current setting range

100 V range: +10.0 A to +42.0 A

(Initial value: +42.0 A)

200 V range: +5.0 A to +21.0 A

(Initial value: +21.0 A)

Negative current setting range:

100 V range:

42.0 A to 10.0 A

(Initial

value:

42.0A)

200 V range:

21.0 A to 5.0 A

(Initial

value:

21.0 A)

Setting resolution:

0.1 A

Output average current limiter
When limiter is operating, suppresses output voltage.

Current setting range

100 V range: 1.0 A to 10.5 A (Initial value: 10.5 A)

200 V range: 1.0 A to 5.3 A

(Initial value: 5.3 A)

Setting resolution:

0.1 A

9.3 Setting Range Limits

When signal source mode is internal or internal + external only

Voltage setting limit
(Lower limit

 upper limit)

Positive voltage setting range

100 V range: +5.0 V to +200.0 V

(Initial value: +200.0 V)

200 V range: +10.0 V to +400.0 V

(Initial value: +400.0 V)

Negative voltage setting range 100 V range:

200.0 V to 5.0 V

(Initial value:

200.0 V)

200 V range:

400.0 V to 10.0 V

(Initial value:

400.0 V)

Setting resolution:

0.1 V

Frequency setting limit
(Lower limit

 upper limit)

Upper limit setting range:

1.0 Hz to 550.0 Hz

(Initial value: 550.0 Hz)

Lower limit setting range:

1.0 Hz to 550.0 Hz

(Initial value: 1.0 Hz)

Setting resolution:

0.1 Hz