9.2 Current Limiter
APS-1102 Programmable AC/DC Power Source
9-6
9.2 Current Limiter
Output peak current limiter
When limiter is operating, output voltage is clipped.
Positive current setting range
100 V range: +10.0 A to +42.0 A
(Initial value: +42.0 A)
200 V range: +5.0 A to +21.0 A
(Initial value: +21.0 A)
Negative current setting range:
100 V range:
42.0 A to 10.0 A
(Initial
value:
42.0A)
200 V range:
21.0 A to 5.0 A
(Initial
value:
21.0 A)
Setting resolution:
0.1 A
Output average current limiter
When limiter is operating, suppresses output voltage.
Current setting range
100 V range: 1.0 A to 10.5 A (Initial value: 10.5 A)
200 V range: 1.0 A to 5.3 A
(Initial value: 5.3 A)
Setting resolution:
0.1 A
9.3 Setting Range Limits
When signal source mode is internal or internal + external only
Voltage setting limit
(Lower limit
upper limit)
Positive voltage setting range
100 V range: +5.0 V to +200.0 V
(Initial value: +200.0 V)
200 V range: +10.0 V to +400.0 V
(Initial value: +400.0 V)
Negative voltage setting range 100 V range:
200.0 V to 5.0 V
(Initial value:
200.0 V)
200 V range:
400.0 V to 10.0 V
(Initial value:
400.0 V)
Setting resolution:
0.1 V
Frequency setting limit
(Lower limit
upper limit)
Upper limit setting range:
1.0 Hz to 550.0 Hz
(Initial value: 550.0 Hz)
Lower limit setting range:
1.0 Hz to 550.0 Hz
(Initial value: 1.0 Hz)
Setting resolution:
0.1 Hz