6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-40

RST
This resets the device to its factory settings. However, reset cannot be executed when output is on.
([1, “Invalid with output on”] error occurs.)

Syntax:

RST

Parameters:

None

Command

example:

RST

Response format:

None

Response example:

None

SAV

This stores the current status information in memory. The memory storage ranges from 1 to 30.

Syntax:

SAV

Parameters:

1 to 30

Command

example:

SAV 1

Response format:

None

Response example:

None

SRE
This sets the service request enable register. The setting ranges from 0 to 178.

Syntax:

SRE

Parameters:

0 to 178

Command

example:

SRE 178

Query

example:

SRE?

Response format:

Response

example: 178

STB?
This returns the status byte register value. The response ranges from 0 to 178.

Syntax:

STB?

Parameters:

None

Query

example:

STB?

Response format:

Response

example: 2