3 grounding and power supply connection – GW Instek APS-1102 User Manual User Manual
Page 25
2.3 Grounding and Power Supply Connection
APS-1102 Programmable AC/DC Power Source
2-5
Do not install the APS-1102 in the following locations:
Location with flammable gas
An explosion may occur. Never install and use this product in such a location.
Outdoors, or location exposed to direct sunlight or near a fire or heat source
The full performance of this product may not be obtained, or failure may occur.
Location with corrosive gas, moisture, or high humidity
This product may become corroded or fail.
Location near an electromagnetic field source, high-voltage device, or power line
This product may malfunction.
Location exposed to excessive vibration
This product may malfunction or fail.
Location with excessive dust
In particular, electrically conductive dust may cause failure of this product.
2.3 Grounding and Power Supply Connection
Be sure to ground the APS-1102.
This product uses a line filter. Be sure to ground this product. Otherwise, an electric shock
may occur.
To prevent an electric shock from occurring, be sure to ground the APS-1102 according to "Electric
Equipment Technical Standard Class D (100
or below) Grounding" or higher.
When the input power line is 100 V AC, use power cord set 1. When a 3-prong power plug that
includes a protective ground contact is connected to a 3-prong power supply outlet, this product is
grounded automatically. Power cord set 1 (Japan version) is rated at 125 V AC.
When the input power line is 200 V AC, use power cord set 2 equipped with a terminal or crimp
contact that suits the type of outlet to be used. Always connect a ground when using power cord set 2.
Power cord set 2 is rated at 250 V AC.
A 3-prong/2-prong conversion adapter is not provided with this product. If a 3-prong/2-prong
conversion adapter is used, the adaptor ground wire must be connected to a ground contact next to the
outlet.
The power requirements of this product are as follows:
Voltage range:
100 V AC to 230 V AC ±10 % (when at 250 V or less)
Overvoltage
category
II
!
WARNING