6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-28

MEASure[:SCALar]:VOLTage[:AC]?
This returns the output voltage RMS value (AC+DC RMS value) in Vrms units. The output
voltage RMS value measurement full scale is 450.0 Vrms. If the measured value exceeds the full
scale, 999.9 Vrms is returned as a measurement error.

Syntax:

MEASure[:SCALar]:VOLTage[:AC]?

Parameters:

None

Query example:

MEAS:VOLT?

Response format:

Response

example: 100.4

MEASure[:SCALar]:VOLTage:HIGH?
This returns the positive voltage peak value (AC+DC peak value) in Vpk units. The positive
voltage peak value measurement full scale is 450 Vpk. If the measured value exceeds the full
scale, 999 Vpk is returned as a measurement error.

Syntax:

MEASure[:SCALar]:VOLTage:HIGH?

Parameters:

None

Query

example:

MEAS:VOLT:HIGH?

Response format:

Response

example: 100

MEASure[:SCALar]:VOLTage:LOW?
This returns the negative voltage peak value (AC+DC peak value) in Vpk units. The negative
voltage peak value measurement full scale is

450 Vpk. If the measured value exceeds the full

scale,

999 Vpk is returned as a measurement error.

Syntax:

MEASure[:SCALar]:VOLTage:LOW?

Parameters:

None

Query

example:

MEAS:VOLT:LOW?

Response format:

Response example:

99