6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-24

d) MEASure subsystem

MEASure[:SCALar]:CURRent[:AC]?
This returns the output current RMS value (AC+DC RMS value) in Arms units. The output
current RMS value measurement full scale is ±15.00 Arms. If the measured value exceeds the full
scale, 99.99 Arms is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent[:AC]?

Parameters:

None

Query example:

MEAS:CURR?

Response format:

Response

example: 12.85

MEASure[:SCALar]:CURRent:HIGH?
This returns the positive current peak value (AC+DC peak value) in Apk units. The positive
current peak value measurement full scale is 45.0 Apk. If the measured value exceeds the full
scale, 99.9 Apk is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent:HIGH?

Parameters:

None

Query

example:

MEAS:CURR:HIGH?

Response format:

Response example:

25.9

MEASure[:SCALar]:CURRent:LOW?
This returns the negative current peak value (AC+DC peak value) in Apk units. The negative
current peak value measurement full scale is

45.0 Apk. If the measured value exceeds the full

scale,

99.9 Arms is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent:LOW?

Parameters:

None

Query

example:

MEAS:CURR:HIGH?

Response format:

Response example:

25.9