GW Instek APS-1102 User Manual User Manual
Page 196
6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source
6-24
d) MEASure subsystem
MEASure[:SCALar]:CURRent[:AC]?
This returns the output current RMS value (AC+DC RMS value) in Arms units. The output
current RMS value measurement full scale is ±15.00 Arms. If the measured value exceeds the full
scale, 99.99 Arms is returned as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent[:AC]?
Parameters:
None
Query example:
MEAS:CURR?
Response format:
Response
example: 12.85
MEASure[:SCALar]:CURRent:HIGH?
This returns the positive current peak value (AC+DC peak value) in Apk units. The positive
current peak value measurement full scale is 45.0 Apk. If the measured value exceeds the full
scale, 99.9 Apk is returned as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent:HIGH?
Parameters:
None
Query
example:
MEAS:CURR:HIGH?
Response format:
Response example:
25.9
MEASure[:SCALar]:CURRent:LOW?
This returns the negative current peak value (AC+DC peak value) in Apk units. The negative
current peak value measurement full scale is
45.0 Apk. If the measured value exceeds the full
scale,
99.9 Arms is returned as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent:LOW?
Parameters:
None
Query
example:
MEAS:CURR:HIGH?
Response format:
Response example:
25.9