6.3 Command Descriptions
APS-1102 Programmable AC/DC Power Source
6-18
[SOURce:]PHASe[:IMMediate]
This sets the phase at the start of output. The phase setting range is 0.0 to 359.9
. However, when
output is on, the phase setting cannot be changed. ([1, “Invalid with output on”] error occurs.)
When in AC-EXT or AC+DC-EXT mode, the output waveform setting cannot be set. ([3,
“Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]PHASe[:IMMediate]
Parameters:
0.0 to 359.9
Command example:
PHAS 90
Query
example:
PHAS?
Response format:
Response
example: 90.0
[SOURce:]SEQuence:CONDition?
This returns the sequence status. The response message is 0 (Idle mode), 1 (Run mode), or 2 (Hold
mode).
Syntax:
[SOURce:]SEQuence:CONDition?
Parameters:
None
Query
example:
SEQ:COND?
Response format:
Response
example: 0
[SOURce:]SEQuence:CSTep?
This returns the step number being executed. The response ranges from
1 and 1 to 255. 1 is
returned in the idle mode.
Syntax:
[SOURce:]SEQuence:CSTep?
Parameters:
None
Query
example:
SEQ:CST?
Response format:
Response
example: 2
[SOURce:]SEQuence:LEN?
This returns the maximum number of sequence steps. 255 is always returned.
Syntax:
[SOURce:]SEQuence:LEN?
Parameters:
None
Query
example:
SEQ:LEN?
Response format:
Response
example: 255