6.6 Error Message List

APS-1102 Programmable AC/DC Power Source

6-49

6.6 Error Message List

USB error messages are listed in Table 6-15 below.

Table 6-15. USB Error Message List

No. Message

Cause

0

“No error”

No error.

100 “Command

error”

Command was recognized as invalid due to any cause.
(General command error)

101 “Invalid

character”

Command was recognized as invalid due to invalid characters
(such as “VOLT&”) in command string.

102 “Syntax

error”

Command was recognized as invalid due to reception of
undefined command or parameter.

200 “Execution

error”

Error occurred when executing the command due to any cause.
(General execution error)

222 “Data out of range”

Error occurred when executing the command due to out of range
parameter.

350 “Too Many Errors”

New errors cannot be retained due to error queue overflow.

1

“Invalid with output on”

Error occurred when executing the command because the output
relay was on.

2

“Invalid with output off”

Error occurred when executing the command because the output
relay was off.

3

“Invalid in this mode”

Error occurred when executing the command due to inappropriate
output mode.

4

“Under Busy State”

Error occurred during execution of command due to busy device.

5

“Out of Limiter”

Error occurred during execution of command because limiter
value was exceeded.

6

“Invalid in this frequency”

Error occurred when executing the command due to inappropriate
oscillation frequency setting.

7

“Out of Input Signal
frequency”

Error occurred when executing the command due to out of the
external synchronization frequency range.

8

“Under Warning State”

Error occurred when executing the command because device was
operating in protected mode.