5.3 SET Menu


APS-1102 Programmable AC/DC Power Source

5-20

5.3.7 Output peak current limiter settings

The positive value and the negative value of the peak current limiter can be set independently.
If the output peak current exceeds the set current, the output voltage is clipped.
Setting range limits can be set for output voltage and output frequency.

See “5.5.5 Setting range limiter (LIMIT)”.

Table 5-16. Output Peak Current Limiter Setting Range

Setting

Setting Range

Resolution

Initial Value

Output peak
current limiter
(positive)

100 V range

+10.0 A to +42.0 A

0.1 A

+42.0 A

200 V range

+5.0 A to +21.0 A

0.1 A

+21.0 A

Output peak
current limiter
(negative)

100 V range

42.0 A to 10.0 A

0.1 A

42.0 A

200 V range

21.0 A to 5.0 A

0.1 A

21.0 A

Limiter settings are retained separately for the 100 V and 200 V range.

Since these are not fully constant current operations, the specified limit value may not be achieved,
depending on the load.

Operation steps

Select the icon in the SET menu, then set a numerical value.
 Output peak current limiter (positive):

 Output peak current limiter (negative):

See “3.4.7 Using limiter functions”, for description of the output peak current limiter

settings.

Note with caution that the output peak current limiter may allow overshooting, depending on the
load, and therefore may not achieve the peak current value as set.
For an inductive load, operation of the limiter can cause an output overvoltage, triggering the
protection function. In this case, the peak current limiter settings (positive and negative) should be
reduced.

If the output current or frequency suddenly increases, due, for example, to an output short circuit,
the limiter response may not be quick enough, so the output overcurrent protection function may be
triggered.

CAUTION

CAUTION