4.5 Control Using External Control I/O Connector

APS-1102 Programmable AC/DC Power Source

4-47

4.5 Control Using External Control I/O Connector

Digital signals that are input to the external control I/O connector can be used to control sequences and the
output on/off setting. The sequence step synchronized output signals are output from the external control
I/O connector.
The status of the APS-1102 can be monitored by reading the output signals from the external control
input/output connector.

See “3.3.4 External signal input/external sync signal input terminal”.

After the external signal input level was changed, approx. 50ms might be required for the controller to
execute a function as a maximum response time. “Output ON/OFF” additionally requires the similar time
for a response of a relay device.
Table 4-13 lists the functions of the connector various pins.

Table 4-13. Functions of External Control I/O Connector

Pin No. Input/Output

Function

1 Output

Outputs power source on/off status.
0 = Off, 1 = On.

2 Output

Outputs the output on/off status.
0 = Off, 1 = On.

3 Output

Outputs the limiter operation status.
0 = Not operating, 1 = Operating.

4 Output

Outputs the software busy status.
0 = Normal operation, 1 = Busy.

5 Output

Step synchronized output 0.
Outputs bit 0 of step synchronized output.

6 Output

Step synchronized output 1.
Outputs bit 1 of step synchronized output.

11 Input

Output off.
Sets output off when falling of input signal is detected.

12 Input

Output on.
Sets output on when falling of input signal is detected.

13 Input

Start sequence.
Starts sequence when falling of input signal is detected.

14 Input

Stop sequence.
Stops sequence when falling of input signal is detected.

15 Input

Hold sequence.
Holds (pauses) sequence when falling of input signal is detected.

16 Input

Sequence branch trigger 0.
Jumps to step specified by BRANCH0 when falling of input signal is detected.

17 Input

Sequence branch trigger 1.
Jumps to step specified by BRANCH1 when falling of input signal is detected.