beautypg.com

GW Instek APS-1102 User Manual User Manual

Page 238

background image

7.3 When Fault Is Suspected

APS-1102 Programmable AC/DC Power Source

7-16

Table 7-13. When Fault Is Suspected (Problem Related to Output Error)

Symptom

Possible Cause

Required Action, etc.

No output

Output voltage setting is “0”.

Check the output voltage setting.

See “3.4.5 Setting output voltage”.

Signal source is not connected.

When the signal source mode is external signal
(EXT), connect the signal source to the external
signal input terminal, then set a suitable external
input gain.

See “3.3.4 External signal

input/external sync signal input
terminal

.

See “4.9 Amplification of External

Signal

.

External input gain is set to “0”.

Output on/off switch
(OUTPUT) is not set to on.

Set the output on/off switch to on.

Protection function-related error
or warning message is
displayed.

Output cannot be set to on when a protection
function-related error or warning message is
being displayed.
See the error responses listed in “Table 7-3.
Protection Function-Related Errors

and

“Table 7-6. Warning Messages” and
eliminate the cause of the error.
When a protection function-related error
message is being displayed, all key operations
become invalid. Turn the power source switch
off, then back on again.

Warning message is
displayed.

Overload has occurred.

Connect a load that is within the maximum
output range or lower the output level.

Excessive signal level in signal
generator

If the signal source is EXT or ADD, either
reduce the connected signal output level or
reduce the external input gain.

Ambient temperature is too
high.

Keep the ambient temperature during use at 40°C
or below

.

Air filter is clogged.

Refer to “8.2 Routine Maintenance” and
clean the air filter.

Something is blocking the air
flow near the air intake on the
front panel or the air exhaust on
the rear panel.

Make sure the APS-1102 complies with the
installation requirements described in “2.2
Installation Environment

.