2.4 Simple Operation Checks

APS-1102 Programmable AC/DC Power Source

2-9

2. Panel settings

Set the APS-1102 as shown below.

See “3.4 Main Operation Example”, for descriptors

of the settings methods.

Table 2-1. Panel Settings for Operation Check

Item Setting

Output mode

AC-INT mode

Output voltage range

100 V

AC output voltage

100 Vrms

Selected measurement display RMS

3. Press

the

OUTPUT

key to turn on output.

OUTPUT

OUTPUT

Output is off.

Output is on.

Check that the control panel MEASURE screen shows 100 Vrms as the AC output voltage.

Figure 2-3. Operation Check

If the normal values are not displayed, it may be due to an operation fault. Contact Good Will

or a Good Will representative.

4. Turn off the power, check that OUTPUT off is displayed, then press the power switch to the bottom

() position. This shuts off the power supply and turns off the power source unit.

Before turning off the power, make sure that the output on/off LED is off.

Output on/off LED is on.

Check display.

CAUTION