6.3 Command Descriptions

APS-1102 Programmable AC/DC Power Source

6-17

[SOURce:]MODE
This sets the output mode. There are eight output modes, each of them is a combination of one
operation mode (AC or AC+DC) and one signal source mode, (INT, EXT, ADD, or SYNC).
Consequently, the output mode setting is AC-INT, AC-EXT, AC-ADD, AC-SYNC, AC+DC-INT,
AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC.
However, when output is on, the output mode cannot be changed. ([1, “Invalid with output on”]
error occurs.)

Syntax:

[SOURce:]MODE

Parameters:

AC-INT/AC-EXT/AC-ADD/AC-SYNC/ACDC-INT

ACDC-EXT/ACDC-ADD/ACDC-SYNC

Command example:

MODE ACDC-INT

Query

example:

MODE?

Response format:

Response

example: ACDC-INT

[SOURce:]PHASe:CLOCk
This sets the sync signal source in the external synchronization mode (AC-SYNC or
AC+DC-SYNC). The sync signal source setting is either LINE (line synchronization) or EXT
(external synchronization). However, when output is on, the sync signal source cannot be changed.
([1, “Invalid with output on”] error occurs.)
Unless the mode is AC-SYNC or AC+DC-SYNC mode, the sync signal source setting cannot be
set. ([3, “Invalid in this mode”] error occurs.)

Syntax:

[SOURce:]PHASe:CLOCk

Parameters:

LINE/EXT

Command example:

PHAS:CLOC EXT

Query

example:

PHAS:CLOC?

Response format:

Response

example: EXT