7.2 Error Messages and Responses

APS-1102 Programmable AC/DC Power Source

7-8

7.2.3 Panel operation errors

Panel operation error messages are displayed in the status window. The message box closes
automatically after 1.5 seconds.

Message box

Figure 7-3. Screen Display When Panel Operation Error Has Occurred

Table 7-4. Panel Operation Errors (1/2)

Error Message

Cause and Action Required

OUT OF MEASURE RANGE

Measurement is out of range, cannot measure.
(

See “3.4.9 Using measurement functions”.)

INVALID IN THIS
FREQUENCY

Frequency setting is not valid, so setting (execution) is disabled. Set
50 Hz or 60 Hz.
(

See “3.4.6 Setting output frequency”.)

OUT OF INPUT SIGNAL
FREQUENCY

External sync signal frequency is outside of sync enabled range.
(

See “4.7 Synchronization of Output to External

Signal”.)

OUT OF RANGE

USB parameter range check error. Set the parameter within the
possible range of settings. (Remote mode only)
(

See “6.3.2 Detailed command descriptions”.)

INVALID WITH OUTPUT ON

Setting (execution) is invalid while output is on. Turn output off.
(

See “3.4.8 Output on/off”.)

INVALID WITH OUTPUT OFF

Setting (execution) is invalid while output is off. Turn output on.
(

See “3.4.8 Output on/off”.)

INVALID IN THIS MODE

Cannot set (execute) due to invalid output mode. Change the output
mode.
(

See “3.4.2 Setting output mode”.)