7.2 Error Messages and Responses
APS-1102 Programmable AC/DC Power Source
7-8
7.2.3 Panel operation errors
Panel operation error messages are displayed in the status window. The message box closes
automatically after 1.5 seconds.
Message box
Figure 7-3. Screen Display When Panel Operation Error Has Occurred
Table 7-4. Panel Operation Errors (1/2)
Error Message
Cause and Action Required
OUT OF MEASURE RANGE
Measurement is out of range, cannot measure.
(
See “3.4.9 Using measurement functions”.)
INVALID IN THIS
FREQUENCY
Frequency setting is not valid, so setting (execution) is disabled. Set
50 Hz or 60 Hz.
(
See “3.4.6 Setting output frequency”.)
OUT OF INPUT SIGNAL
FREQUENCY
External sync signal frequency is outside of sync enabled range.
(
See “4.7 Synchronization of Output to External
Signal”.)
OUT OF RANGE
USB parameter range check error. Set the parameter within the
possible range of settings. (Remote mode only)
(
See “6.3.2 Detailed command descriptions”.)
INVALID WITH OUTPUT ON
Setting (execution) is invalid while output is on. Turn output off.
(
See “3.4.8 Output on/off”.)
INVALID WITH OUTPUT OFF
Setting (execution) is invalid while output is off. Turn output on.
(
See “3.4.8 Output on/off”.)
INVALID IN THIS MODE
Cannot set (execute) due to invalid output mode. Change the output
mode.
(
See “3.4.2 Setting output mode”.)