GW Instek GDS-3000 Series GKT-100 deskew fixture User Manual User Manual

Deskew fixture

background image

Deskew Fixture

GKT-100

USER MANUAL

ISO-9001 CERTIFIED MANUFACTURER

G

ETTING STARTED

The GKT-100 deskew fixture is used to compensate for

the propagation delay between a passive voltage probe

and current probe. It is used with the GDS-3000.

Required tools

• GDS-3000 x 1
• GKT-100 x 1
• USB type A-B cable x1 -used for deskew fixture
• Standard passive probe x1
• Current probe x1 (GCP-530 or GCP-1030)

C

ONNECT THE DESKEW FIXTURE

Before performing the manual deskew procedure,

please connect the deskew fixture to the USB host port

on the GDS-3000. Connect a passive voltage probe to

the JP1 terminal on the fixture and the current probe as

shown in the next picture.

1

3

2

4

Item Description

1

Connect the USB cable: Connect the USB cable to the

USB port on the deskew fixture and the other end of

the cable to the USB host port on the GDS-3000.

2

Connect the current probe: Connect a current probe

(GCP-530 or GCP-1030) to the loop on the deskew

fixture and the other end of the current probe to the

CH1 terminal on the front panel of the GDS-3000.

3

Connect the passive probe: Connect the passive

probe to the JP2 terminal on the deskew fixture and

the other end to the CH2 terminal on the GDS-3000

front panel.

4

Grounding: Connect the grounding clip of the passive

voltage probe to JP1 terminal on the deskew fixture.

Note • Demagnetize and zero the current probe. Refer

to the current probe’s documentation for

instructions on how to do this.

Make sure the direction of the arrow on the

current probe and that on the current loop are

pointing in the same direction.

P

ERFORMING DESKEW PROCEDURE

Follow the deskew procedure steps below once the

deskew fixture and probes are connected.

Step

1. Press the Autoset key on the

front panel of GDS-3000.

2. Press the CH1 key.

3. Press Probe from the bottom

menu.

4. Press Deskew on the side menu

and use the variable knob to set

the deskew time for the current

probe.

This manual contains proprietary information, which is

protected by copyright. All rights are reserved. No part

of this manual may be photocopied, reproduced or

translated to another language without prior written

consent of Good Will Corporation.
The information in this manual was correct at the time

of printing. However, Good Will continues to improve

its products and therefore reserves the right to change

the specifications, equipment, and maintenance

procedures at any time without notice.

Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng
Dist., New Taipei City 236, Taiwan.

5. Press the CH2 key.

6. Repeat steps 3 to step 4 to adjust

the deskew time for the passive

voltage probe.

7. Adjust the waveform output of

voltage and current so that no

lead or lag is shown between

both waveforms to finish the

deskew adjustment.

S

PECIFICATIONS

The specifications apply when the oscilloscope is

powered on for at least 30 minutes under +20°C~+30°C.

Model Specific Specifications

Temperature

Ambient temperature: -10 ~ 55°C

Storage Temperature: 20 ~60°C, no

condensation

Relative humidity: 93% @40°C

65% @ 41°C~60°C
Storage: 90% @ 65 °C for 24 hr

Altitude

Operating: to 4,570 m (15,000ft)
Storage: to 15,244 m (50,000 ft)

Humidity

GW INSTEK PART NO. 82KT-10000M01