Display commands, Cursor:xy:ratio:delta, Display:intensity:waveform – GW Instek GDS-2000A series Programming Manual User Manual
Page 73

COMMAND DETAILS
73
Parameter
1, 2 (Cursor 1, Cursor 2)
Return parameter
Example
:CURSor:XY:RATio:POSition?
6.717E+1
Returns the ratio value as 6.717V/V.
:CURSor:XY:RATio:DELta
Description
Queries the ratio delta in XY mode.
Syntax
:CURSor:XY:RATio:DELta {?}
Return parameter
Example
:CURSor:XY:RATio:DELta?
5.39E+1
Returns the ratio delta as 53.9V/V.
Display Commands
:DISPlay:INTensity:WAVEform ................................... 73
:DISPlay:INTensity:GRATicule.................................... 74
:DISPlay:PERSistence ................................................. 74
:DISPlay:GRATicule .................................................... 74
:DISPlay:WAVEform .................................................... 76
:DISPlay:OUTPut ........................................................ 76
:DISPlay:INTensity:WAVEform
Description
Sets or queries the waveform intensity level.
Syntax
:DISPlay:INTensity:WAVEform {
Parameter
0.0E+0 to 1.0E+2 (0% to 100%)
Return Parameter
Query
Set
Query