Trigger:bus:b1:i2c:address:value, Trigger:bus:b1:i2c:address:direction – GW Instek GDS-2000A series Programming Manual User Manual

Page 137

background image

COMMAND DETAILS

137

Return Parameter

Returns the address type

Example

:TRIGger:BUS:B1:I2C:ADDRess:TYPe?
CBUS

:TRIGger:BUS:B1:I2C:ADDRess:VALue

Description

Sets or queries the I

2

C bus address value when the

I

2

C bus is set to trigger on Address or

Address/Data.

Syntax

:TRIGger:BUS:B1:I2C:ADDRess:VALue {string | ? }

Related
commands

:TRIGger:BUS:B1:I2C:ADDRess:MODe

Parameter

7/10 characters, must be enclosed in
double quotes, "string".

x = don’t care

1 = binary 1

0 = binary 0

Return Parameter

Returns the address value.

Example1

:TRIGger:BUS:B1:I2C:ADDRess:VALue "xxx0101"
Sets the address to XXX0101

Example 2

:TRIGger:BUS:B1:I2C:ADDRess:VALue?
XXX0101

:TRIGger:BUS:B1:I2C:ADDRess:DIRection

Description

Sets or queries the address bit as read write or
don’t care.

Note

This setting only applies when the I

2

C trigger is set

to trigger on Address or Address/Data

Syntax

:TRIGger:BUS:B1:I2C:ADDRess:DIRection { READ |
WRITE | NOCARE | ? }

Set

Query

Set

Query