Trigger:bus:b1:lin:identifier:value – GW Instek GDS-2000A series Programming Manual User Manual

Page 159

background image

COMMAND DETAILS

159

:TRIGger:BUS:B1:LIN:IDentifier:VALue

Description

Sets or returns the binary address string to be used
for the LIN trigger.
Note: Only applicable when the condition is set to
DATA or IDANDDATA.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).

Syntax

:TRIGger:BUS:B1:LIN:IDentifier:VALue {|?}

Parameter/
Return parameter

The size of the string depends on the
data size setting. The string must be
enclosed in double quotes, “string”.

String contents:
x = don’t care
1 = binary 1
0 = binary 0

Example

:TRIGger:BUS:B1:LIN:CONDition ID
:TRIGger:BUS:B1:LIN:IDentifier:VALue "00X1X01X"
:TRIGger:BUS:B1:LIN:IDentifier:VALue?
>01100X1X01X

Set