beautypg.com

Trigger:bus:b1:lin:errtype – GW Instek GDS-2000A series Programming Manual User Manual

Page 158

background image

GDS-2000A Programming Manual

158

Parameter/
Return parameter

The size of the string depends on the
data size setting. The string must be
enclosed in double quotes, “string”.

String contents:
x = don’t care
1 = binary 1
0 = binary 0

Example

:TRIGger:BUS:B1:LIN:DATa:SIZe 1
:TRIGger:BUS:B1:LIN:DATa:VALue "01010X1X"
:TRIGger:BUS:B1:LIN:DATa:VALue?
>01010X1X

:TRIGger:BUS:B1:LIN:ERRTYPE

Description

Sets or returns the error type be used for the LIN
trigger.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).

Syntax

:TRIGger:BUS:B1:LIN:ERRTYPE
{SYNC|PARIty|CHecksum|?}

Parameter/
Return parameter

SYNC

Sets the LIN error type to SYNC.

PARIty

Sets the LIN error type to parity.

CHecksum

Sets the LIN error type to checksum.

Example

:TRIGger:BUS:B1:LIN:ERRTYPE?
>SYNC
:TRIGger:BUS:B1:LIN:ERRTYPE CHECKSUM
:TRIGger:BUS:B1:LIN:ERRTYPE?
>CHECKSUM

Set