COMMAND DETAILS

141

Return Parameter

Returns the data value.

Example1

:TRIGger:BUS:B1:UART:CONDition RXDATA
:TRIGger:BUS:B1:UART:RX:DATa:SIZe 1
:TRIGger:BUS:B1:UART:RX:DATa:VALue "1x1x0101"
Sets the value to 1x1x0101

Example 2

:TRIGger:BUS:B1:UART:RX:DATa:VALue?
1X1X0101

:TRIGger:BUS:B1:UART:TX:DATa:SIZe

Description

Sets or queries the number of bytes for UART data.

Note

This setting only applies when the UART trigger is
set to trigger on Tx Data

Syntax

:TRIGger:BUS:B1:UART:TX:DATa:SIZe { | ?}

Related
commands

:TRIGger:BUS:B1:UART:CONDition

Parameter

Number of bytes (1 to 10).

Return parameter

Returns the number of bytes.

Example

:TRIGger:BUS:B1:UART:TX:DATa:SIZe 5
Sets the number of bytes to 5.

:TRIGger:BUS:B1:UART:TX:DATa:VALue

Description

Sets or queries the triggering data value for the
UART bus when the bus is set to trigger on Tx
Data.

Syntax

:TRIGger:BUS:B1:UART:TX:DATa:VALue {string | ? }

Related
commands

:TRIGger:BUS:B1:UART:TX:DATa:SIZe

Set

Query

Set