beautypg.com

Measurement:meas:maximum – GW Instek GDS-2000A series Programming Manual User Manual

Page 106

background image

GDS-2000A Programming Manual

106

Example

:MEASUrement:MEAS1:SOUrce1 CH1
:MEASUrement:MEAS1:TYPe PK2PK
:MEASUrement:MEAS1:STATE ON
:MEASUrement:MEAS1:VALue?
5.000E+0
Selects channel 1 as the source for measurement 1,
sets measurement 1 to peak to peak measurement
and then turns on the measurement. The result
returns the peak to peak measurement.

:MEASUrement:MEAS:MAXimum

Description

Returns the maximum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.

Syntax

:MEASUrement:MEAS:MAXimum?

Related
Commands

:MEASUrement:STATIstics:MODe

Parameter

MEAS The automatic measurement number

from 1 to 8.

Example

:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MAXimum?
2.800E-02
Returns the maximum measurement result for
measurement number 3.

Query