COMMAND DETAILS

157

Example

:TRIGger:BUS:B1:LIN:DATa:QUALifier?
>EQUAL
:TRIGger:BUS:B1:LIN:DATa:QUALifier THAN
:TRIGger:BUS:B1:LIN:DATa:QUALifier?
>THAN

:TRIGger:BUS:B1:LIN:DATa:SIZe

Description

Sets or returns the length of the data string in bytes
for the LIN trigger.
Note: Only applicable when the condition is set to
DATA or IDANDDATA.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).

Syntax

:TRIGger:BUS:B1:LIN:DATa:SIZe {|?}

Parameter/
Return parameter

1~8 (bytes)

Example

:TRIGger:BUS:B1:LIN:DATa:SIZe?
>1
:TRIGger:BUS:B1:LIN:DATa:SIZe 2
:TRIGger:BUS:B1:LIN:DATa:SIZe?
>2

:TRIGger:BUS:B1:LIN:DATa:VALue

Description

Sets or returns the binary data string to be used for
the LIN trigger.
Note: Only applicable when the condition is set to
DATA or IDANDDATA.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).

Related
Commands

:TRIGger:BUS:B1:LIN:DATa:SIZe

Syntax

:TRIGger:BUS:B1:LIN:DATa:VALue {|?}

Set

Query

Set