Trigger:bus:b1:can:identifier:mode, Trigger:bus:b1:can:identifier:value – GW Instek GDS-2000A series Programming Manual User Manual
Page 151
COMMAND DETAILS
151
Return parameter REMote
Sets the frame type to remote frame
ERRor
Sets the frame type to error frame
OVERLoad
Sets the frame type to overload
Example
:TRIGger:BUS:B1:CAN:FRAMEtype DATA
Sets the frame type to DATA.
:TRIGger:BUS:B1:CAN:IDentifier:MODe
Description
Sets or returns the CAN addressing mode for the
bus.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
Syntax
:TRIGger:BUS:B1:CAN:IDentifier:MODe
{STANDard|EXTended|?}
Parameter/
Return parameter
STANDard
Standard addressing mode
EXTended
Extended addressing mode
Example
:TRIGger:BUS:B1:CAN:IDentifier:MODe?
>STANDARD
Returns the addressing mode.
:TRIGger:BUS:B1:CAN:IDentifier:VALue
Description
Sets or returns the binary address string used for
the CAN trigger.
Note: Only applicable when the trigger condition
is set to ID or IDANDDATA.
Requirements: Firmware version V1.19 or later with
DS2-LA8 (or DS2-LA16).
Syntax
:TRIGger:BUS:B1:CAN:IDentifier:VALue {
Related
Commands
:TRIGger:BUS:B1:CAN:IDentifier:MODe
Set
Query
Set
Query