Timebase:scale, Timebase:mode, Timebase:window:position – GW Instek GDS-2000A series Programming Manual User Manual

Page 116

background image

GDS-2000A Programming Manual

116

:TIMebase:SCALe

Description

Sets or queries the horizontal scale.

Syntax

:TIMebase:SCALe { | ?}

Parameter

Horizontal scale

Return parameter

Returns the horizontal scale.

Example

:TIMebase:SCALe 5.00E-2

Sets the horizontal scale to 50ms/div.

:TIMebase:MODe

Description

Sets or queries the time base mode. The time base
mode determines the display view window on the
scope.

Syntax

:TIMebase:MODe {MAIN | WINDow | XY | ?}

Parameter

MAIN

Sets the time base mode to the main
screen.

WINDow

Sets the time base mode to the zoom
window.

XY

Sets the time base mode to the XY
display.

Return parameter

Returns the time base mode (MAIN, WINDOW,
XY)

Example

:TIMebase:MODe MAIN

Sets the time base mode to the main mode.

:TIMebase:WINDow:POSition

Description

Sets or queries the zoom horizontal position.

Syntax

:TIMebase:WINDow:POSition { | ?}

Set

Query

Set

Query

Set

Query