beautypg.com

Segments:save:select:start, Segments:save:select:end – GW Instek GDS-2000A series Programming Manual User Manual

Page 247

background image

COMMAND DETAILS

247

:SEGMents:SAVe:SELect:STARt

Description

Sets or queries the starting segment to save from.

Syntax

:SEGMents:SAVe:SELect:STARt
{SETTOMIN | SETTOMAX | | ? }

Related
commands

:SEGMents:TOTalnum

Parameter/
Return parameter

SETTOMIN

Sets the starting segment to min
segment.

SETTOMAX

Sets the starting segment to the max
segment.

Sets the segment to 1~2048.

Example

:SEGMents:SAVe:SELect:STARt 2
Sets the starting segment to segment number 2.

:SEGMents:SAVe:SELect:END

Description

Sets or queries the ending segment to save from.

Syntax

:SEGMents:SAVe:SELect:END
{SETTOMIN | SETTOMAX | | ? }

Related
commands

:SEGMents:TOTalnum

Parameter/
Return parameter

SETTOMIN

Sets the starting segment to min
segment.

SETTOMAX

Sets the starting segment to the max
segment.

Sets the segment to 1~2048.

Example

:SEGMents:SAVe:SELect:END 10
Sets the ending segment to segment number 10.


Set

Query

Set