COMMAND DETAILS

243

TABLE

Sets the measurement mode to a
measurement list.

Example1

:SEGMents:MEASure:MODe?
>PLOT
Returns the measurement mode as Statistics.

:SEGMents:MEASure:PLOT:SOURce

Description

Sets or queries the statistics source.

Syntax

:SEGMents:MEASure:PLOT:SOURce { | ? |}

Related
commands

:SEGMents:MEASure:MODe ; :SEGMents:MEASure:P
LOT:DIVide ; :SEGMents:MEASure:PLOT:SELect ; :SE
GMents:MEASure:PLOT:RESults

Parameter/
Return parameter

1~8 (Automatic measurement item
1~8)

Example1

:SEGMents:MEASure:PLOT:SOURce 1
Sets the source as auto measurement item 1.

:SEGMents:MEASure:PLOT:DIVide

Description

Sets or queries the number of bins for the statistics
function.

Syntax

:SEGMents:MEASure:PLOT:DIVide { | ? |}

Related
commands

:SEGMents:MEASure:PLOT:SOURce ; :SEGMents:ME
ASure:PLOT:SELect

Parameter/
Return parameter

1~20

Example1

:SEGMents:MEASure:PLOT:DIVide 5
Sets the number bins to 5 for the statistics function.

Set

Query

Set

Query