GDS-2000A Programming Manual

162

:RECAll:WAVEform

Description

Recalls a waveform from wave1~wave20 or from
file to REF1~4.

Note

Only *.LSF files can be recalled using this
command. *.CSV files cannot be recalled.

Syntax

:RECAll:WAVEform{W |
(“Disk:/xxx.LSF”,”USB:/xxx.LSF”)},REF

Parameter

n

1~20 (Wave1~wave20)

xxx.LSF

Filename in file path.

1,2,3,4 (REF1, REF2, REF3, REF4)

Example

:RECAll:WAVEform W1, REF1

Recalls the waveform stored in Wave1 to reference
1.

:SAVe:IMAGe

Description

Saves a screen image to the assigned file path with
a specified filename.

Syntax

:SAVe:IMAGe { (“Disk:/xxx.PNG”,
“USB:/xxx.BMP)}

Related
commands

:SAVe:IMAGe:FILEFormat
:SAVe:IMAGe:INKSaver

Parameter

xxx.PNG or
xxx.BMP

File name (8 characters max)

Example

:SAVe:IMAGe “Disk:/pic1.PNG”
Saves a screen image named pic1.png to the root
directory (Disk:/) of the scope.
:SAVe:IMAGe “USB:/pic1.BMP”
Saves a screen image named pic1.bmp to the root
directory of the external USB flash disk.

Set