beautypg.com

Measurement:meas:stddev, Measurement:statistics:mode – GW Instek GDS-2000A series Programming Manual User Manual

Page 108

background image

GDS-2000A Programming Manual

108

Example

:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MINImum?
1.600E-02
Returns the minimum measurement result for
measurement number 3.

:MEASUrement:MEAS:STDdev

Description

Returns the standard deviation for the selected
measurement from the last time the statistics were
reset. This is a statistics related command.

Syntax

:MEASUrement:MEAS:STDdev?

Related
Commands

:MEASUrement:STATIstics:MODe

Parameter

MEAS The automatic measurement number

from 1 to 8.

Example

:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:MEAS3:STATE ON
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:STDdev?
1.530E-03
Returns the standard deviation for measurement
number 3.

:MEASUrement:STATIstics:MODe

Description

Puts the statics measurement results on the display
or queries whether the statics are displayed.

Syntax

:MEASUrement:STATIstics:MODe {OFF | ON | ?}

Query

Set

Query