beautypg.com

Chapter 3 operating hapsite in portable mode, 1 starting the hapsite in portable mode, Chapter 3 – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 95

background image

3 - 1

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Chapter 3

Operating HAPSITE in Portable Mode

3.1 Starting the HAPSITE in Portable Mode

Portable Mode refers to using the HAPSITE without the Laptop computer.

Required Materials

HAPSITE (Analytical Module)

Internal Standard Gas Canister

Carrier Gas Canister

Charged Battery

AC to DC Power Converter Power Supply

Probe

Procedure

1

Assemble the HAPSITE as shown in

Section 2.2, Basic Assembly, on page 2-5

.

2

Press the POWER button on the front panel. The word POWER will illuminate.
Powering on the HAPSITE takes one (1) to two (2) minutes. See

Figure 3-1

.

Figure 3-1 The Power Switch

NOTE: Power on the HAPSITE while connected to AC power. Using battery

power to turn on and heat the HAPSITE will consume over 40% of the
battery’s charge.

3

The HAPSITE will boot in approximately one minute and will sense which
sample configuration (i.e., concentrator) has been installed. It will begin to
prepare the default method for this sample configuration.

4

The HAPSITE will begin to prepare various components. These components
include heating the HAPSITE and accessory heaters, running AutoTune (see
Step

8 on page 3-3

), powering the NEG, and if necessary, running concentrator

cleanout.