16 - 31

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

4

Remove the new NEG Pump from the protective container. See

Figure

16-29

.

CAUTION

The NEG Pump is stored under a vacuum. The shelf life
is 5 years. If the vacuum is compromised, shelf life is
reduced to 1 year.

5

Remove the o-ring and seating ring assembly from the new NEG Pump. See

Figure 16-29

.

Figure 16-29 New NEG Pump

6

On the new NEG Pump, wipe inside the flange with methanol on a lint free
wipe. See

Figure 16-30

.

Figure 16-30 Cleaning Flange with Methanol

O-ring and Seating

Assembly