HAPSITE Smart Plus Operating Manual

Figure 9-14 Using the Range Tool

9.5.3.1 Steps Required to Range Acquisition

1

Place the cursor on the triangle with Two Red Vertical Lines (R1). Click and
hold the LMB. Then, drag R1 to the location where the scans should be
averaged. Refer to

Figure 9-14

.

2

Move the cursor to the tip of the R1 marker. The cursor should change to a
vertical double-headed arrow. Holding the LMB, moving the double-headed
arrow up widens a range. Moving the arrow down narrows the range of the
peak averaging. The red range lines should intersect the peak sides at 50% of
their height. The left and right side boundaries can be manually adjusted using
a drag and drop with the LMB.