HAPSITE Smart Plus Operating Manual

5.7.2 Quick Reference SOP — Analyze Mode with SituProbe

1

Double-click the Plus IQ software icon.

2

Verify that the desired sample configuration (i.e., Sample Loop or
Concentrator) is installed.

3

Make sure the SituProbe is set-up properly. See

Section 2.9, SituProbe, on

page 2-27

for assembly information.

4

If necessary, power on the HAPSITE or take the system out of Extended
Standby.

5

Depending upon the method, the HAPSITE will begin preparing the
Concentrator. If needed, see

Section 3.1.2, Concentrator Options (CONC

OPTIONS), on page 3-8

. See

Section 5.3, Selecting a New Method, on page

5-7

if the start up method is not the desired method

6

When the system has finished preparing, click the RUN METHOD icon. Select
the desired method from the appropriate folder. Refer to Steps 2-4 of

section

5.3 on page 5-7

.

7

Click the Run button in the pop-up window.

8

When the screen prompts, Collect Sample Now and Sampling Is Done, no
action is required from the user.

9

When the run is complete, a Method Finished prompt will appear.

10

To analyze the data, see

Chapter 9, Data Review

.