HAPSITE Smart Plus Operating Manual

5.4.1 Quick Reference SOP — Running Analyze (GC/MS) Loop Method

1

Verify that the Sample Loop is installed.

2

Power on or take the HAPSITE out of Extended Standby.

3

The HAPSITE will begin to prepare the start up Loop method. See

Section 5.3,

Selecting a New Method, on page 5-7

if the start up method is not the desired

method.

4

Double-click the Plus IQ icon to open the software.

5

If the HAPSITE runs an AutoTune, press OK when the screen prompts.

6

Click the RUN METHOD icon and choose the method that has been prepared.
Refer to Steps 2-4 of

section 5.3 on page 5-7

.

7

Click the Run button in the pop-up window of the software

8

When the screen prompts, Collect Sample Now, hold the probe over the
sample for the specified amount of time.

9

Remove the Probe from the sample when the screen prompts, Sampling Is
Done
.

10

Let the method run to completion.

11

Review data when run is complete.