11.5.7 SituProbe Flow Parameter
WARNING
Only trained users should modify methods. Incorrect
flow rates may lead to improper analysis.
SituProbe Flow Pressure controls the flow rate of Nitrogen through the SituProbe
during the Sample and Purge states. This parameter is only available when the
SituProbe is used.
11.6 Tune Page
The Tune Page is divided into two pages, Report and Param, that each provide
information about the Tune file.
11.6.1 Param Page
The Param Page (see
Figure 11-21
) provides the filename of the Tune file on the
HAPSITE, which provides the MS tune parameters used with the method. These
parameters are updated after every successful tune. The default filename is
default.tun
. If a different Tune file exists on the HAPSITE then the Browse
button can be used to locate and specify the desired tune for the Method. The
default.tun
file always contains the most recent Tune parameters.
This page also has a Show Details checkbox (
Figure 11-22
), which will produce a
grid of Tune parameters contained in the file and produced by the Tune Report.
These parameters cannot be edited.
11.6.2 Report Page
The Report Page (
Figure 11-23
) provides a printable report saved after the most
recent execution of AutoTune.